Chapter 13 217
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:CALC{1-4}:DATA:SDAT
Syntax :CALCulate{[1]|2|3|4}[:SELected]:DATA:SDATa <numeric 1>,…,<numeric NOP×2>
:CALCulate{[1]|2|3|4}[:SELected]:DATA:SDATa?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4) (specified with the
:CALC{1-4}:PAR{1-4}:SEL command), sets/reads out the corrected data array. (It is the
data array for which processing such as error correction to measured raw data has been
performed. For details, refer to “Corrected data arrays” on page 108.)
The data transfer format when this command is executed depends on the setting with the
:FORM:DATA command.
Query response {numeric 1},…,{numeric NOP×2}<newline><^END>
Where NOP is the number of measurement points and n is an integer between 1 and NOP.
Example of use
10 DIM A(1:201,1:2)
20 OUTPUT 717;":CALC1:DATA:SDAT?"
30 ENTER 717;A(*)
Related commands :CALC{1-4}:DATA:SMEM on page 218
:CALC{1-4}:DATA:FDAT on page 215
:FORM:DATA on page 338
Equivalent key No equivalent key is available on the front panel.
Description
{numeric n×2-1} Real part of the data (complex number) at the n-th measurement
point.
{numeric n×2} Imaginary part of the data (complex number) at the n-th
measurement point.