316 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:DISP:TABL
Syntax :DISPlay:TABLe[:STATe] {ON|OFF|1|0}
:DISPlay:TABLe[:STATe]?
Description Turns ON/OFF the display of the window selected with the :DISP:TABL:TYPE
command.
Parameters
Query response {1|0}<newline><^END>
Example of use
10 OUTPUT 717;":DISP:TABL ON"
20 OUTPUT 717;":DISP:TABL?"
30 ENTER 717;A
Related commands :DISP:TABL:TYPE on page 317
Equivalent key
[Sweep Setup] - Edit Segment Table
[Marker] - Marker Table
[Analysis] - Limit Test - Edit Limit Line
[Macro Setup] - Echo Window
NOTE When performing the operation from the front panel, you select the window and turn
ON/OFF the display at the same time.
Description
ON or 1 Turns ON the display.
OFF or 0 (preset value) Turns OFF the display.