Chapter 13 397
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SENS{1-4}:CORR:COLL:ECAL:SOLT1
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:ECAL:SOLT1 <numeric>
Description Executes full 1-port calibration of the specified port of channel 1 (:SENS1) to channel 4
(:SENS4) using the ECal (Electrical Calibration) module.
If you execute this command when the ECal module is not connected, an error occurs and
the command is ignored. (No query)
Parameters
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:ECAL:SOLT1 1"
20 OUTPUT 717;"*OPC?"
30 ENTER 717;A
Equivalent key [Cal] - ECal - 1-Port Cal - Port 1|Port 2
:SENS{1-4}:CORR:COLL:ECAL:SOLT2
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:ECAL:SOLT2 <numeric 1>,<numeric 2>
Description Executes full 2-port calibration between the 2 specified ports of channel 1 (:SENS1) to
channel 4 (:SENS4) using the ECal (Electrical Calibration) module.
If you execute this command when the ECal module is not connected, an error occurs and
the command is ignored. (No query)
Parameters
For each parameter, you must specify a different port number. If you specify the same port
number for 2 or more parameters, an error occurs and the command is ignored.
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:ECAL:SOLT2 1,2"
20 OUTPUT 717;"*OPC?"
30 ENTER 717;A
Equivalent key [Cal] - ECal - 2-Port Cal
<numeric>
Description Port number
Range 1 to 2
Resolution 1
<numeric 1> <numeric 2>
Description Port number Port number
Range 1 to 2 1 to 2
Resolution 1 1