Agilent Technologies E5071C Guide de l'utilisateur Page 418

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 551
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 417
418 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:FREQ:STAR
Syntax :SENSe{[1]|2|3|4}:FREQuency:STARt <numeric>
:SENSe{[1]|2|3|4}:FREQuency:STARt?
Description Sets the start value of the sweep range of channel 1 (:SENS1) to channel 4 (:SENS4).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:FREQ:STAR 100E6"
20 OUTPUT 717;":SENS1:FREQ:STAR?"
30 ENTER 717;A
Related commands :SENS{1-4}:FREQ:STOP on page 419
Equivalent key
[Start]
<numeric>
Description Start value
Range 3E5 to 3E9
Preset value 3E5
Unit Hz (hertz)
Resolution 1
Vue de la page 417
1 2 ... 413 414 415 416 417 418 419 420 421 422 423 ... 550 551

Commentaires sur ces manuels

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm