Agilent Technologies E5071C Guide de l'utilisateur Page 300

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 551
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 299
300 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:CONT:HAND:E
Syntax :CONTrol:HANDler:E[:DATA] <numeric>
:CONTrol:HANDler:E[:DATA]?
Description When input/output port E (port C + port D) of the handler I/O is set to the output port,
outputs data to output port E.
When input/output port E is set to the input port, reads out data inputted to port E.
Data is outputted as 8-bit binary using C0 as LSB and D3 as MSB.
For details about the handler I/O, see Chapter 10.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":CONT:HAND:C:MODE OUTP"
20 OUTPUT 717;":CONT:HAND:D:MODE OUTP"
30 OUTPUT 717;":CONT:HAND:E 128"
10 OUTPUT 717;":CONT:HAND:C:MODE INP"
20 OUTPUT 717;":CONT:HAND:D:MODE INP"
30 OUTPUT 717;":CONT:HAND:E?"
40 ENTER 717;A
Related commands :CONT:HAND:C:MODE on page 297
:CONT:HAND:D:MODE on page 299
:CONT:HAND:C on page 296
:CONT:HAND:D on page 298
Equivalent key No equivalent key is available on the front panel.
<numeric>
Description Output data/input data
Range 0 to 255
Resolution 1
Vue de la page 299