Chapter 13 373
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SENS{1-4}:BWID
Syntax :SENSe{[1]|2|3|4}:BWIDth[:RESolution] <numeric>
:SENSe{[1]|2|3|4}:BWIDth[:RESolution]?
Description Sets the IF bandwidth of channel 1 (:SENS1) to channel 4 (:SENS4).
This command provides the same function as the :SENS{1-4}:BAND command.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:BWID 1E3"
20 OUTPUT 717;":SENS1:BWID?"
30 ENTER 717;A
Related commands :SENS{1-4}:BAND on page 372
Equivalent key
[Avg] - IF Bandwidth
:SENS{1-4}:CORR:CLE
Syntax :SENSe{[1]|2|3|4}:CORRection:CLEar
Description Clears all callibration coefficient and measured standard data for calibration in the channel
1 (SENS 1) to channel 4 (SENS 4).
Example of use
10 OUTPUT 717;":SENS1:CORR:CLE"
Related commands
Equivalent key
[Cal] - Clear All
<numeric>
Description IF bandwidth
Range 10 to 30000
Preset value 30000
Unit Hz (hertz)
Resolution In steps of 1 or 3