Agilent Technologies E5071C Guide de l'utilisateur Page 417

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 551
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 416
Chapter 13 417
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SENS{1-4}:FREQ:SPAN
Syntax :SENSe{[1]|2|3|4}:FREQuency:SPAN <numeric>
:SENSe{[1]|2|3|4}:FREQuency:SPAN?
Description Sets the span value of the sweep range of channel 1 (:SENS1) to channel 4 (:SENS4).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:FREQ:SPAN 1E9"
20 OUTPUT 717;":SENS1:FREQ:SPAN?"
30 ENTER 717;A
Related commands :SENS{1-4}:FREQ:CENT on page 415
Equivalent key
[Span]
<numeric>
Description Span value
Range 0 to 2.9997E9
Preset value 2.9997E9
Unit Hz (hertz)
Resolution 1
Vue de la page 416
1 2 ... 412 413 414 415 416 417 418 419 420 421 422 ... 550 551

Commentaires sur ces manuels

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm