380 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:CORR:COLL:CKIT:ORD:THRU
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:ORDer:THRU <numeric 1>,<numeric
2>, <numeric 3>
:SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:ORDer:THRU? <numeric 1>,<numeric
2>
Description For the calibration kit selected for channel 1 (:SENS1) to channel 4 (:SENS4), selects the
standard used for the thru measurement between the specified 2 ports.
Parameters
For <numeric 1> and <numeric 2>, you must specify a different port number. If you
specify the same port number, an error occurs and the command is ignored.
If the specified parameter is out of the allowable setup range, an error occurs and the
command is ignored.
Query response {numeric 3}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:CKIT:ORD:THRU 1,2,11"
20 OUTPUT 717;":SENS1:CORR:COLL:CKIT:ORD:THRU? 1,2"
30 ENTER 717;A
Related commands :SENS{1-4}:CORR:COLL:CKIT on page 375
Equivalent key
[Cal] - Modify Cal Kit - Specify CLSs - Thru
:SENS{1-4}:CORR:COLL:CKIT:RES
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:RESet
Description Resets the calibration kit selected for channel 1 (:SENS1) to channel 4 (:SENS4) to the
factory setting state. (No query)
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:CKIT:RES"
Related commands :SENS{1-4}:CORR:COLL:CKIT on page 375
Equivalent key
[Cal] - Modify Cal Kit - Restore Cal Kit
<numeric 1> <numeric 2> <numeric 3>
Description Port number 1 Port number 2 Standard number
Range 1 to 2 1 to 2 1 to 21
Resolution 1 1 1