Agilent Technologies InfiniiVision 3000 DSO-X 3054A Manuel de service Page 140

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 970
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 139
140 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
5 Common (*) Commands
*RST (Reset)
(see page 860)
Command Syntax
*RST
The *RST command places the instrument in a known state. This is the
same as pressing [Save/Recall] > Default/Erase > Factory Default on the front
panel.
When you perform a factory default setup, there are no user settings that
remain unchanged. To perform the equivalent of the front panel's [Default
Setup] key, where some user settings (like preferences) remain unchanged,
use the :SYSTem:PRESet command.
Reset conditions are:
Acquire Menu
Mode Normal
Averaging Off
# Averages 8
Analog Channel Menu
Channel 1 On
Channel 2 Off
Volts/division 5.00 V
Offset 0.00
Coupling DC
Probe attenuation AutoProbe (if AutoProbe is connected),
otherwise 1.0:1
Vernier Off
Invert Off
BW limit Off
Impedance 1 M Ohm
Units Volts
Skew 0
Cursor Menu
Source Channel 1
Vue de la page 139
1 2 ... 135 136 137 138 139 140 141 142 143 144 145 ... 969 970

Commentaires sur ces manuels

Pas de commentaire