482 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
24 :SBUS<n> Commands
:SBUS<n>:I2S:SOURce:DATA
(see page 860)
Command Syntax
:SBUS<n>:I2S:SOURce:DATA <source>
<source> ::= {CHANnel<n> | EXTernal} for the DSO models
<source> ::= {CHANnel<n> | DIGital<d>} for the MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :SBUS<n>:I2S:SOURce:DATA command controls which signal is used as
the serial data (SDATA) source by the serial decoder and/or trigger when
in I2S mode.
Query Syntax
:SBUS<n>:I2S:SOURce:DATA?
The :SBUS<n>:I2S:SOURce:DATA? query returns the current source for the
I2S serial data (SDATA).
Return Format
<source><NL>
See Also "Introduction to :TRIGger Commands" on page 641
":SBUS<n>:I2S:SOURce:CLOCk" on page 481
":SBUS<n>:I2S:SOURce:WSELect" on page 483