216 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
9 :CALibrate Commands
:CALibrate:PROTected
(see page 860)
Query Syntax
:CALibrate:PROTected?
The :CALibrate:PROTected? query returns the rear-panel calibration
protect (CAL PROTECT) button state. The value PROTected indicates
calibration is disabled, and UNPRotected indicates calibration is enabled.
Return Format
<switch><NL>
<switch> ::= {PROT | UNPR}
See Also "Introduction to :CALibrate Commands" on page 211