:SEARch:SERial:CAN:PATTern:DATA
(see page 860)
Command Syntax
:SEARch:SERial:CAN:PATTern:DATA <string>
<string> ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X}
for hexadecimal
The :SEARch:SERial:CAN:PATTern:DATA command specifies the data value
when searching for Data Frame ID and Data.
The length of the data value is specified using the
:SEARch:SERial:CAN:PATTern:DATA:LENGth command.
Query Syntax
:SEARch:SERial:CAN:PATTern:DATA?
The :SEARch:SERial:CAN:PATTern:DATA? query returns the current data
value setting.
Return Format
<string><NL>
<string> ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X}
for hexadecimal
See Also Chapter 25, “:SEARch Commands,” starting on page 557
":SEARch:SERial:CAN:MODE" on page 587