Agilent Technologies InfiniiVision 3000 DSO-X 3054A Manuel de service Page 553

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 970
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 552
:SBUS<n> Commands 24
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 553
:SBUS<n>:UART:TRIGger:IDLE
(see page 860)
Command Syntax
:SBUS<n>:UART:TRIGger:IDLE <time_value>
<time_value> ::= time from 1 us to 10 s in NR3 format
The :SBUS<n>:UART:TRIGger:IDLE command selects the value of the idle
period for burst trigger in the range from 1 us to 10 s when in UART
mode.
Query Syntax
:SBUS<n>:UART:TRIGger:IDLE?
The :SBUS<n>:UART:TRIGger:IDLE? query returns the current UART trigger
idle period time.
Return Format
<time_value><NL>
<time_value> ::= time from 1 us to 10 s in NR3 format
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:MODE" on page 648
":SBUS<n>:UART:TRIGger:BURSt" on page 551
":SBUS<n>:UART:TRIGger:TYPE" on page 555
Vue de la page 552
1 2 ... 548 549 550 551 552 553 554 555 556 557 558 ... 969 970

Commentaires sur ces manuels

Pas de commentaire