Agilent B1500A/B1505A Programming Guide, Edition 11 2-19
Remote Mode Functions
Measurement Modes
• When the quasi-pulse source applies voltage close to the stop value.
• When the quasi-pulse source reaches its current compliance due to the
breakdown condition of the device under test.
NOTE If the slew rate was too slow when settling detection started or if the settling
detection time was too long, an error occurs and the source returns its output to the
start value immediately. See “BDM” on page 4-45.
4. After the settling detection stops, the quasi-pulse source keeps the output.
5. After the delay time, the measurement channel starts measurement.
Only one channel can be used for measurement.
6. After measurement, the quasi-pulse source immediately returns the output to the
start value and keeps it.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
NOTE If there is noise or skew on the output voltage, settling detection might stop at an
unexpected voltage.
NOTE The BDT command sets the hold time and delay time, and the BDM command sets
the settling detection interval and measurement mode (voltage or current); the BDV
command sets the output. Also |start-stop| must be 10 V or more.