Agilent Technologies B1500A Manuel de service Page 125

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 598
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 124
Agilent B1500A/B1505A Programming Guide, Edition 11 2-39
Remote Mode Functions
Measurement Modes
NOTE The WTFC command sets the hold time, delay time, and step delay time.
The WMFC command sets the automatic abort function and the output after
measurement.
The WFC command sets the AC signal frequency sweep output.
The ACV command specifies the oscillator level, and applies the AC signal.
The DCV command applies the specified DC bias.
If the SCUU (SMU CMU Unify Unit) is connected to the CMU and two
MP/HRSMU modules correctly, the source module is automatically selected by the
DC bias setting. The CMU is selected if it is ±25 V or less (setting resolution: 0.001
V), or the SMU is selected if it is greater than ±25 V (setting resolution: 0.005 V).
To Stop Sweep
Output
An automatic abort function is available. Refer to “Automatic Abort Function” on
page 2-46.
Vue de la page 124
1 2 ... 120 121 122 123 124 125 126 127 128 129 130 ... 597 598

Commentaires sur ces manuels

Pas de commentaire