Agilent Technologies B1500A Manuel de service Page 393

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 598
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 392
Agilent B1500A/B1505A Programming Guide, Edition 11 4-77
Command Reference
DZ
If the output voltage is greater than the allowable voltage for the interlock open
condition, the interlock circuit must be shorted.
Syntax DV chnum,vrange,voltage[,Icomp[,comp_polarity[,irange]]
Parameters chnum : SMU source channel number. Integer expression. 1 to 10 or 101 to
1001. See Table 4-1 on page 4-16.
vrange : Ranging type for voltage output. Integer expression. The output range
will be set to the minimum range that covers voltage value. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See Table 4-4 on page 4-21.
voltage : Output voltage value (in V). Numeric expression. See Table 4-7 on
page 4-24, Table 4-9 on page 4-26, Table 4-12 on page 4-27, or Table
4-15 on page 4-28 for each measurement resource type.
Icomp : Current compliance value (in A). Numeric expression. See Table 4-7 on
page 4-24, Table 4-9 on page 4-26, Table 4-12 on page 4-27, or Table
4-15 on page 4-28 for each measurement resource type. If you do not
set Icomp, the previous value is used. 0 A is not allowed for Icomp.
comp_
polarity : Polarity of current compliance. Integer expression.
0: Auto mode (default setting). The compliance polarity is
automatically set to the same polarity as voltage, regardless of the
specified Icomp. If voltage=0 V, the polarity is set to positive.
1: Manual mode. Uses the polarity of Icomp you specified.
irange : Current compliance ranging type. Integer expression. The compliance
range will be set to the minimum range that covers Icomp value. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See Table 4-5 on page 4-22.
Example
Statements
OUTPUT @B1500;"DV 1,0,20,1E-6,0,15"
OUTPUT @B1500;"DV 2,12,10"
DZ
The DZ command stores the settings (V/I output values, V/I output ranges, V/I
compliance values, and so on) of the specified channels, and sets the channels to 0
V. The settings can be recovered by using the RZ command. The stored settings are
cleared by using a device clear (HP BASIC CLEAR) command, *RST, RZ, CL, CA,
or *TST?.
Vue de la page 392
1 2 ... 388 389 390 391 392 393 394 395 396 397 398 ... 597 598

Commentaires sur ces manuels

Pas de commentaire