Agilent B1500A/B1505A Programming Guide, Edition 11 4-103
Command Reference
ERSSP?
ERSSP?
This query command returns the connection state of the I/O path for the Agilent
16440A SMU/PGU selector (B1500A-A04).
Syntax ERSSP? port
Parameters port: Output port of SMU/PG selector. Integer expression.
0: Output 1 on selector of first module
1: Output 2 on selector of first module
2: Output 1 on selector of second module
3: Output 2 on selector of second module
Query Response status<CR/LF^EOI>
0: Open. Normally open mechanical relay contact.
1: SMU connect. Connected to the SMU input.
2: PGU connect. Connected to the PGU input.
3: PGU open. Open semiconductor relay contact installed on the PGU connect path.
Example
Statements
OUTPUT @B1500;"ERSSP? 1"
ENTER @B1500;A
See Also “ERMOD”
ERUHVA
This command specifies the MC/HCSMU modules connected to the V Control and I
Control inputs of Agilent N1268A ultra high voltage expander.
Execution
Conditions
No SMU may be in the high voltage state (forcing more than the allowable voltage
for the interlock open condition, or voltage compliance set to more than it).
The CL command has been already executed for the measurement resources
specified in the ERUHVA command.
Digital I/O port must be set to the N1268A control mode using the ERMOD 16
command.
Syntax ERUHVA vsmu,ismu