4-108 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
IN
TC command
TTC command
Spot C measurement (MM17)
CV (DC bias) sweep measurement (MM18)
Pulsed spot C measurement (MM19)
Pulsed sweep CV measurement (MM20)
C-f sweep measurement (MM22)
CV (AC level) sweep measurement (MM23)
C-t sampling measurement (MM26)
Execution
Conditions
This command is not effective for the binary data output format (FMT3, FMT4,
FMT13, and FMT14). Then one of the following couples will be measured. They
will be automatically selected by the B1500, and will be a couple without data
overflow.
R (resistance, Ω) and X (reactance, Ω)
G (conductance, S) and B (susceptance, S)
Syntax IMP mode
Parameters mode : Measurement mode. 1 to 402. Integer expression. See Table 4-16 on
page 4-29.
Initial setting: mode=100 (Cp-G measurement)
Example
Statements
OUTPUT @B1500;"IMP 10"
IN
The IN command sets the specified channel to 0 V with an output range change.
Syntax IN [chnum[,chnum...[,chnum]...]]
A maximum of 15 channels can be set.
Parameters chnum : Channel number. Integer expression. See Table 4-1 on page 4-16.
If you specify multiple chnums, the channel outputs will be set to 0 V in the
specified order.