5-26 Agilent B1500A/B1505A Programming Guide, Edition 11
Error Messages
Operation Error
5309 CRC for N1265A EEPROM Vm correction segment failed.
N1265A might be defective. Contact your nearest Agilent Technologies
service center.
5310 CRC for N1265A EEPROM Im correction segment failed.
N1265A might be defective. Contact your nearest Agilent Technologies
service center.
5311 Specified module is already assigned to voltage control or current
control of HVMCU.
Specify a free MCSMU/HCSMU.
5312 Different module must be assigned to voltage control and current
control of HVMCU.
Specify a free MCSMU/HCSMU.
5313 Do not execute CN/CNX to the current control module set to
ERHVCA.
Execute the CN/CNX command to the module that is set to voltage
control by the ERHVCA command.
5314 HVSMU Current Expander output must be open state.
Open the input-output path before switching the HVSMU series
resistor.
5315 Execute CL to the module set to ERHVCA.
Execute the CL command to the modules that will be specified by the
ERHVCA command.
5316 CRC for N1266A EEPROM Vm correction segment failed.
N1266A might be defective. Contact your nearest Agilent Technologies
service center.
5317 CRC for N1266A EEPROM Im correction segment failed.
N1266A might be defective. Contact your nearest Agilent Technologies
service center.
5318 Over voltage is detected in Selector Output High Force, Low Force, or
Low Sense terminal of N1265A.
Remove the causes of overvoltage.