1-48 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Basics
Data Output Format
8 Bytes Data Elements
To use the 8 bytes binary data format, enter the FMT13 or FMT14 command.
The data (Data, Source_data, Sampling_no, Data_search, Data_sense, Osc_level,
and Dc_bias) will be sent as the binary value shown in Figure 1-3. The format of the
time data (Time) will be different from the others.
Figure 1-3 8 Bytes Binary Data Output Format
A: Type. One bit.
B: Parameter. Seven bits.
C: Range. One byte. Range value used to calculate the data.
D: Data count. Four bytes.
E: Status. One byte.
F: Channel number. Five bits.
G: A/D converter. Three bits.
H: Time data count. Six bytes.
These data elements are described in the following pages.
65432107 65432107 65432107 65432107
Byte 5
Byte 6
Byte 7
Byte 8
65432107 65432107 65432107 65432107
Byte 1
Byte 2
Byte 3
Byte 4
A
B
C
D
E
G
F
65432107 65432107 65432107 65432107
Byte 5 Byte 6 Byte 7 Byte 8
65432107 65432107 65432107 65432107
Byte 1 Byte 2 Byte 3 Byte 4
A
B
H
F
For time data:
For measurement data and source data: