4-182 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
SPV
Syntax SPUPD chnum[,chnum...[,chnum[,chnum]]...]
A maximum of ten channels can be set.
Parameters chnum : SPGU channel number. Integer expression. 1 to 10 or 101 to 1002.
If multiple chnums are specified, all outputs are started in the specified order. The
channel numbers 1 to 10 correspond to the channel numbers 101 to 1001
respectively. See Table 4-1 on page 4-16.
Example
Statements
OUTPUT @B1500;"CN 101,102,201,202"
OUTPUT @B1500;"SPUPD 101,102,201,202"
OUTPUT @B1500;"SRP"
SPV
This command sets the output voltage of the specified SPGU channel. Set for each
channel. For the parameters, see Figure 2-31 on page 2-57.
Execution
Conditions
The SPGU operating mode must be set to PG with the SIM 0 command.
Syntax SPV chnum,src,base[,peak]
Parameters chnum : SPGU channel number. Integer expression. 1 to 10 or 101 to 1002. See
Table 4-1.
src : Channel signal source. Integer expression. Initial setting 1.
0: DC bias source
1: Pulse signal source 1
2: Pulse signal source 2
base : Pulse base voltage or DC output voltage. Numeric expression. -40 V to
40 V, setting resolution 1 mV. Initial setting 0 V.
peak : Pulse peak voltage. Numeric expression. -40 V to 40 V, setting
resolution 1 mV. Initial setting 0 V.
If no value is set for peak, the base value will be used for both
parameters.
The peak parameter has no effect on DC bias sources.
Example
Statements
OUTPUT @B1500;"SPV 101,1,-0.5,0.5"