1-16 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Basics
Getting Started
To Disable Source/Measurement Channels
To disable the channels, send the CL command. The B1500 opens the output switch
of the specified channels. Opening the output switch disables the channel.
Example OUTPUT @B1500;"CL 1"
This example disables channel 1 (the module installed in slot 1 of the B1500). If you
do not specify the channel, the CL command disables all channels.
To Control ASU
This function is available for Agilent B1500A. ASU (atto sense and switch unit) has
two inputs, SMU input for the B1511B/B1517A and AUX input for the other
instrument. And the ASU input to output connection can be controlled by the
following commands. When the B1500 is turned on, the SMU input will be
connected to the ASU output. However, the SMU output switch will be off at this
time.
Table 1-4 ASU Input Output Connection Control
where, chnum must be the channel number of the SMU connected to the ASU. See
“SAL”, “SAP”, and “SAR” on page 4-172 for the other function and control
commands of the ASU.
When the SMU side is connected to the ASU output, the source output on/off can be
controlled by the CN/CL command. And then the SAP chnum, 1 command is used
to change the output connection to the AUX side. When the AUX side is connected,
the output of the instrument connected to the AUX input is appeared to the ASU
output immediately.
Previous Connection Command Subsequent Connection
SMU side, Output on/off SAP chnum, 1 AUX side
SMU side, Output off CN chnum SMU side, Output on
SAP chnum, 0
AUX side CN chnum
SAP chnum, 0
CL [chnum] SMU side, Output off
SMU side, Output on CL [chnum]