2-86 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Trigger Function
Trig In/Out Internal Circuit
The following figures show the trigger input/output circuits internally connected to
the Trig In/Out connectors.
Figure 2-51 Trigger Input Internal Circuit
Figure 2-52 Trigger Output Internal Circuit
Vcc
R1
R2
to trigger
input control
from Trig In (BNC)
Vcc=5 V
R1=220 ohm
R2=10 kohm
C1=1000 pF
C1
Vcc
R1
from trigger
output control
to Trig Out (BNC)
Vcc=5 V
R1=150 ohm
C1=1000 pF
74ABT245 or
equivalent
C1