4-84 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
ERHPR?
Parameters pin : Relay control output pin number. Integer expression. 1 to 6.
state : Logical state. Integer expression. 0 or 1.
0: OFF (0 V, circuit common), initial setting
1: ON (12 V)
Example
Statements
OUTPUT @B1500;"ERHPR 1,1"
OUTPUT @B1500;"ERHPR 2,1"
ERHPR?
Only for the N1258A module selector users. This command returns the logical state
set to the specified relay control output pin.
Syntax ERHPR? pin
Parameters pin : Relay control output pin number. Integer expression. 1 to 6.
Query Response state<CR/LF^EOI>
0 (logical state OFF, 0 V) or 1 (logical state ON, 12 V). Integer expression.
Example
Statements
OUTPUT @B1500;"ERHPR? 1"
ENTER @B1500;A
ERHPS
The ERHPS command sets the connection status for the series resistor on the
HVSMU path of Agilent N1258A/N1259A module selector.
Execution
Conditions
Digital I/O port must be set to the N1258A/N1259A control mode using the
ERMOD 2 command.
Syntax ERHPS onoff
Parameters onoff : 1 (connect) or 0 (disconnect, initial setting). Integer expression.
Example
Statements
OUTPUT @B1500;"ERHPS 1"