Agilent Technologies B1500A Manuel de service Page 123

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 598
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 122
Agilent B1500A/B1505A Programming Guide, Edition 11 2-37
Remote Mode Functions
Measurement Modes
NOTE The PTDCV command sets the pulse timing parameters, such as pulse width and
pulse period.
The WMDCV command sets the automatic abort function.
The PWDCV command sets the pulsed bias sweep voltage. The base, start, and stop
values must have the same polarity for log sweep.
The FC command sets the AC signal frequency.
The ACV command specifies the oscillator level, and applies the AC signal.
If the SCUU (SMU CMU Unify Unit) is connected to the CMU and two
MP/HRSMU modules correctly, the source module is automatically selected by the
sweep range (from start to stop). The CMU is selected if it is ±25 V or less (setting
resolution: 0.001 V), or the SMU is selected if it is greater than ±25 V (setting
resolution: 0.005 V).
To Stop Sweep
Output
An automatic abort function is available. Refer to “Automatic Abort Function” on
page 2-46.
Vue de la page 122
1 2 ... 118 119 120 121 122 123 124 125 126 127 128 ... 597 598

Commentaires sur ces manuels

Pas de commentaire