3-118 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Examples
Using Trigger Function
Using Trigger Function
The Agilent B1500 can be equipped with eight trigger ports that will be used for
different purpose individually. The Agilent B1500 can synchronize the operation
with other equipment by using the trigger function. For details about the trigger
input/output operation, see “Trigger Function” on page 2-74. The following
commands are available for the trigger function.
Command Function and Syntax
TGP Sets the trigger port for the specified terminal.
TGP port,terminal,polarity[,type]
TGPC Clears the trigger setting of the specified ports.
TGPC [port ... [,port] ... ]
TGSI Selects the sweep step first or last that ignores the Start Step
Output Setup trigger input set by the TGP port,1,polarity,2
command.
TGSI mode
TGSO Selects the trigger type, edge or gate, for the Step Output Setup
Completion trigger output set by the TGP port,2,polarity,2
command.
TGSO mode
TGXO Selects the trigger type, edge or gate, for the Measurement
Completion trigger output set by the TGP port,2,polarity,1
command.
TGXO mode
TGMO Selects the trigger type, edge or gate, for the Step Measurement
Completion trigger output set by the TGP port,2,polarity,3
command.
TGMO mode
TM3 Enables the trigger set by the TGP port,terminal,polarity,1
command.