4-222 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
WNCC
2: Stop value.
If this parameter is not set, the MFCMU forces the start value.
Output Data The B1500 returns the data measured before an abort condition is detected. Dummy
data 199.999E+99 will be returned for the data after abort.
Example
Statements
OUTPUT @B1500;"WMFC 2"
OUTPUT @B1500;"WMFC 2,2"
WNCC
The WNCC command clears the multi channel sweep setup. This command is
effective for the measurement modes 16, 27, and 28, and clears the setup of the
following commands.
WNX, MCPT, MCPNT, MCPNX, MCPWS, and MCPWNX
Syntax WNCC
Example
Statement
OUTPUT @B1500;"WNCC"
WNU?
The WNU? query command returns the number of sweep steps specified by the
sweep command (WI, WV, PWI or PWV), and stores the results in the output data
buffer (query buffer).
Execution
Conditions
If you want to know the number of steps for a pulsed sweep, you must execute an
“MM 4” command before using this command, otherwise the number of steps for
the staircase sweep is reported.
Syntax WNU?
Query Response number_of_sweep_steps<CR/LF^EOI>
Example
Statement
OUTPUT @B1500;"WNU?"
ENTER @B1500;A
WNX
The WNX command specifies the staircase sweep source (synchronous sweep
source) that will be synchronized with the primary sweep source. Including the
primary sweep source, the maximum of ten sweep sources can be used for a