Agilent Technologies EasyEXPERT Guide de l'utilisateur Page 163

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 228
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 162
Agilent EasyEXPERT Users Guide Vol. 2, Edition 1 10-13
Error Message
Agilent EasyEXPERT Operation Error
105987 N1265A Self-calibration failed. Code: A, Reason: B
Confirm the connection of the control modules and perform the
self-calibration again.
105988 N1266A Self-test failed. Code: A, Reason: B
Confirm the connection of the control modules and perform the
self-calibration of the modules. Then try the self-test again.
105989 Cannot use the N1258A/N1265As series resistor for HVSMU when
the HVMCU is used.
105990 N1268A Ultra High Voltage Expander is not found.
105991 Communication with N1268A Ultra High Voltage Expander
interrupted. Connect the cable properly.
105992 N1268A Ultra High Voltage Expander is turned off. Connect the power
cable properly and check the power switch.
105993 Same SMU cannot be specified for V Control SMU and I Control SMU
of N1268A Ultra High Voltage Expander.
105994 A is already used in B.
105995 Diagnosis failed. Code: A, Reason: B
Confirm the connections of modules and the connections of output
terminals. Then try the diagnosis again.
105996 Diagnosis failed. Code: A, Reason: B
Confirm the connections of modules and the open condition of output
terminals. Then try the diagnosis again.
105998 Main frame reset occurs.
105999 Digital I/O devices are turned off. Connect the power cable properly
and check the power switch.
106001 Logon request was rejected. The number of Logons already reaches the
limit of the software regulation. Restart EasyEXPERT. If this does not
solve the problem, restart the B1500.
107001 Workspace management error was occurred.
107002 Workspace was not found. Restart Agilent B1500. If the problem still
occurs, contact your nearest Agilent Technologies.
107003 Workspace could not be loaded. Restart Agilent B1500. If the problem
still occurs, contact your nearest Agilent Technologies.