Agilent EasyEXPERT Users Guide Vol. 2, Edition 1 8-15
Application Library
QSCV Maximum Measurement Value and Accuracy
QSCV Maximum Measurement Value and
Accuracy
This section describes the maximum measurement value and accuracy for the result
data measured by using the QSCV[2] application test. These values do not guarantee
the measurement performance. They are just the supplemental information.
“Maximum Measurement Value”
“Measurement Accuracy”
NOTE To obtain stable measurement results, use two source monitor units (SMU). Assign
the current meter and the voltage sweep source to the individual SMU.
Maximum Measurement Value
NOTE The maximum measurement value is not the specifications but the supplemental
data.
Even if the capacitance is within this range, measurement may not be completed due
to influences on the measurement path such as extension cables, etc. The maximum
measurement value will be smaller depending on the leakage current.
The maximum measurement value depends on the settings of the current
measurement range, the QSCV measurement voltage, and the integration time. See
Figure 8-1 to Figure 8-5. Each figure shows the characteristics of the capacitance
value vs. the QSCV measurement voltage by the integration time setting. Each line
in the graph specifies the maximum measurement value for each integration time
setting.
The graph data is applied to the measurement when the QSCV integration time is set
to the same value as the LEAK integration time.