10-32 Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1
Error Message
Agilent FLEX Execution Error
Agilent FLEX Execution Error
When Agilent B1500 causes errors in Agilent FLEX command execution
environment, the B1500 returns the following error code and error message.
100 Undefined GPIB command.
Send the correct command.
102 Incorrect numeric data syntax.
Correct the data syntax.
103 Incorrect terminator position.
Correct the command syntax. The number of parameters will be
incorrect.
104 Incorrect serial data syntax.
120 Incorrect parameter value.
Correct the parameter value.
121 Channel number must be 1 to 10.
Correct the channel number. The channel number must be 1 to 10 for
Agilent B1500.
122 Number of channels must be corrected.
Check the MM, FL, CN, CL, IN, DZ, or RZ command, and correct the
number of channels.
123 Compliance must be set correctly.
Incorrect compliance value was set. Set the compliance value correctly.
124 Incorrect range value for this channel.
Check the range value available for the channel, and correct the range
value.
125 Search goal value must be less than compliance value.
126 Pulse base and peak must be same polarity.
The polarity of the base and peak values must be the same in the PI
command. Also the polarity of the base, start, and stop values must be
the same in the PWI command.