Agilent EasyEXPERT Users Guide Vol. 2, Edition 1 10-35
Error Message
Agilent FLEX Execution Error
205 DZ must be sent before RZ.
The RZ command is effective for the channels set to 0 V output by the
DZ command.
206 Do not specify the channel recovered by RZ.
Specify the channels that have not been recovered yet by the RZ
command after the DZ command. The RZ command cannot be
executed if the specified channels include a channel that has already
been recovered by the RZ command.
210 Ext trigger could not start measurement.
External trigger cannot start measurement because of busy condition.
211 TM1 must be sent to use GET.
Send the TM1 command to use the GPIB GET command (TRIGGER
statement in HP BASIC).
212 Compliance must be set correctly.
Compliance was not set or an incorrect compliance value was set in the
DV, DI, PV, PI, PWV, PWI, TDV, TDI, LSV, LSI, LSSV, LSSI, BSV,
BSI, BSSV, or BSSI command. Set the compliance value correctly.
213 Cannot perform self-test or calibration.
Self-test and calibration cannot be performed in the high voltage state.
Set the output voltage or the voltage compliance to the allowable
voltage for the interlock open condition or less to perform the self-test
or calibration.
214 Send MM before measurement trigger.
Before sending the measurement trigger, the MM command must be
sent to set the measurement mode.
217 Self-test is not defined for this module.
218 Cannot change output range.
The specified module failed the output range change. The output is set
to the initial condition 0 V and 100 A.
219 Channel output switch must be OFF.
The channel output switch must be OFF before executing the specified
command.