Agilent Technologies EasyEXPERT Guide de l'utilisateur Page 97

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 228
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 96
Agilent EasyEXPERT Users Guide Vol. 2, Edition 1 8-21
Application Library
QSCV Maximum Measurement Value and Accuracy
Table 8-3 Constant Value for Calculating the Measurement Accuracy: MPSMU,
HRSMU, ASU+SMU
Measurement
Range Constant
Voltage Output Range
QSCV Operating
Mode
0.5 V 2 V 20 V 40 V 100 V 200 V
1 nA Ap 0.11 0.11 0.11 0.11 0.12 0.12
Bp 0.014 0.029 0.051 0.153 0.298 0.720 Normal
0.009 0.011 0.015 0.033 0.056 0.131 4155C/4156C
compatible
Cp 0.026
Dp 0.014
Ao 0.03 Normal
0.04 4155C/4156C
compatible
Bo 1.7E-15
Boc 1.3E-04
Co 5.9E-14
Coc 4.5E-04
Do 1.7E-15
Doc 1.3E-04
10 nA Ap 0.11 0.11 0.11 0.11 0.12 0.12
Bp 0.014 0.029 0.051 0.153 0.298 0.720 Normal
0.009 0.011 0.015 0.033 0.056 0.131 4155C/4156C
compatible
Cp 0.031
Dp 0.015
Ao 0.03 Normal
0.04 4155C/4156C
compatible
Bo 1.7E-15
Boc 7.3E-05
Co 3.9E-14
Coc 1.9E-04
Do 1.7E-15
Doc 7.3E-05
Vue de la page 96