10-46 Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1
Error Message
Agilent FLEX Execution Error
1002 Illegal pulse width
Check the pulse width and set the correct value.
1003 Illegal pulse base/peak limit
Check the pulse base and peak values, and set the correct values.
1004 Illegal pulse base/peak polarity
Check the pulse base and peak values, and set the polarity properly.
1005 Illegal sweep polarity
Check the sweep start and stop values, and set the polarity properly.
1006 Application measurement setup is not sufficient.
Check the setup required for the specified measurement and set it
properly.
1007 Source channel must be set.
Set the source output channel properly.
1008 Pulse output channel is required.
Specify the pulse output channel. Or set the pulse output channel
properly.
1010 This B1500A does not support PCH.
PCH is not supported by this B1500A. The B1500A must be upgraded
to use the dual HCSMU operation.
1011 SMU is not installed in the slot set to PCH.
Empty slot or non SMU module was specified by PCH. The HCSMU
modules must be specified.
1012 Execute CL to the modules set to PCH.
Output switch of the modules set to PCH must be OFF previously.
1013 Execute ERHPA 0,0,0 before PCH.
Modules set to ERHPA cannot be specified in PCH.
Execute
ERHPA 0,0,0, then execute PCH. After that execute ERHPA to
set the module selector input connection.
1014 Modules set to PCH must be HCSMU.
Only the HCSMU module can be specified in PCH.