10-56 Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1
Error Message
Agilent FLEX Execution Error
5319 Over voltage is detected in Selector Output High Sense terminal of
N1265A.
Remove the causes of overvoltage.
5320 Pulse width overrun is detected in N1265A.
Check the assignment of control modules.
5321 Specified module is already assigned to voltage control or current
control of UHVU.
Specify a free MCSMU/HCSMU.
5322 Different module must be assigned to voltage control and current
control of UHVU.
Specify a free MCSMU/HCSMU.
5323 Do not execute CN/CNX to the current control module set to
ERUHVA.
Execute the CN/CNX command to the module that is set to voltage
control by the ERUHVA command.
5324 Execute CL to the module set to ERUHVA.
Execute the CL command to the modules that will be specified by the
ERUHVA command.
5325 Specified module is already used for UHCU, Selector, or Gate of
N1265A.
Specify a free module.
5326 Specified module is already used for HVMCU.
Specify a free module.
5327 Specified module is already used for UHVU.
Specify a free module.
5330 N1265A: Cooling fan is not operating.
Check the operation of cooling fan and the condition around the fan.
5340 N1265A: Control modules are not assigned. Execute ERPFUHCA to
assign control modules.
5341 N1265A: Status test failed.
5342 N1265A: Voltage pulse test failed.