Agilent EasyEXPERT Users Guide Vol. 2, Edition 1 10-41
Error Message
Agilent FLEX Execution Error
320 Excess current in CMU.
Current that exceeds maximum current at the present voltage range was
detected by the CMU. The output switch was set to OFF.
321 This command is not available for CMU.
CMU was specified for the SMU dedicated command. Specify SMU.
322 This command is not available for SMU.
SMU was specified for the CMU dedicated command. Specify CMU.
323 Use SSP instead of CN for SCUU modules.
It is not necessary to specify the modules connected to the SMU CMU
Unify Unit (SCUU) in the CN command. The output switches will be
controlled by the SSP command.
330 Turn on again to detect source channel.
SCUU might be disconnected/connected on the power on condition.
Restart the B1500 to detect and enable the channel.
331 Turn on again to detect synchronous channel.
SCUU might be disconnected/connected on the power on condition.
Restart the B1500 to detect and enable the channel.
332 Turn on again to detect measurement channel.
SCUU might be disconnected/connected on the power on condition.
Restart the B1500 to detect and enable the channel.
333 Turn on again to detect search source.
SCUU might be disconnected/connected on the power on condition.
Restart the B1500 to detect and enable the channel.
334 Turn on again to detect search sync source.
SCUU might be disconnected/connected on the power on condition.
Restart the B1500 to detect and enable the channel.
335 Turn on again to detect search monitor channel.
SCUU might be disconnected/connected on the power on condition.
Restart the B1500 to detect and enable the channel.
603 Sweep and pulse channels must be different.
Set the sweep source and the pulse source to different channels for the
staircase sweep with pulsed bias measurement (MM5).