2- 56 Agilent B1500 User’s Guide, Edition 7
Introduction
Specifications
HV-SPGU
HV-SPGU: High Voltage Semiconductor Pulse Generator Unit
Number of output channels: 2 channels per module
Modes: pulse, constant, and free run
Standard pulse mode
Two level pulse
Three level pulse per one channel
Pulse period: 30 ns to 10 s
Delay range: 0 s to 9.99 s
Delay resolution: minimum 2.5 ns
Output count: 1 to 1,000,000
Voltage monitor minimum sampling period: 5 μs
Trigger output
Level: TTL
Timing: Synchronized with pulse period
Trigger width: Pulse period × 1
/2 (pulse
period 10 μs), maximum 5 μs (pulse
period > 10 μs)
Pulse range and pulse parameter: See Table 2-29.
Pulse/DC output voltage and accuracy: See Table 2-30.
Figure 2-9 SPGU pulse setup parameters
Tperiod
Twidth
Tdelay
Vamp
Vpeak
Vbase
Tr Tf
90%
90%
10%
10%