Agilent B1500 User’s Guide, Edition 7 2- 59
Introduction
Specifications
• Programmable timing range: 10 ns to 10 s, 10 ns resolution
• SPGU supplemental information
• Pulse width jitter: 0.001 % + 150 ps
• Pulse period jitter: 0.001 % + 150 ps
• Maximum slew rate: 1000 V/μs (
50 Ω load
)
• Noise: 10 mVrms (at DC output)
• Advanced feature:
Voltage monitor: The HV-SPGU has a voltage monitor function to measure
th
e voltage at
the DUT terminal.
Measurement accuracy (open load): ± (0.1 % of reading + 25 mV)
Measurement resolution: 50 μV
Note: Specified at 1 PLC (20 ms = (5 μs s
ample + 5
μs interval)× 2000
samples)
Voltage compensation: The HV-SPGU can measure the imped
ance of DUT
and adjust the output voltage according to the DUT impedance.
Figure 2-10 ALWG setup table and pattern example
Figure 2-11 ALWG complex waveform example
Point Time Voltage
1 0 0.0 V
2 50 ns 0.0 V
3 70 ns 15.0 V
4 100 ns 15.0 V
5 200 ns 0.0 V
6 300 ns 0.0 V
7 320 ns 10.0 V
8 400 ns 10.0 V
9 450 ns 0.0 V
10 500 ns 0.0 V
P1
P4P3
P7 P8
P9P6P5P2
P10
0
500 ns
15
10
0
Pattern
Pattern1 x 1 Pattern2 x 5
Pattern3 x 1