2- 26 Agilent B1500 User’s Guide, Edition 7
Introduction
Measurement Units
Table 2-11 HRSMU Measurement Voltage Values and Resolutions
Table 2-12 HRSMU Measurement Current Values and Resolutions
Range
Measurement
Value
a
Measurement Resolutions
High Speed ADC High Resolution ADC
0.5 V 0 |V| 0.55 V 25 μV 0.5 μV
2 V 0 |V| 2.2 V 100 μV 2 μV
5 V 0 |V| 5.5 V 250 μV 5 μV
20 V 0 |V| 22 V 1 mV 20 μV
40 V 0 |V| 44 V 2 mV 40 μV
100 V 0 |V| 100 V 5 mV 100 μV
Range
Measurement
Value
a
a. This column is applied to the auto ranging or the limited auto ranging.
For fixed ranging, maximum measurement value is Range column value.
Measurement Resolutions
High Speed ADC
High Resolution ADC
1 pA
b
b. Available when the E5288A Atto Sense and Switch Unit (ASU) is used.
0 |I| 1.15 pA 100 aA 100 aA
10 pA 0 |I| 11.5 pA 1 fA
1 fA / 400 aA
b
100 pA 0 |I| 115 pA 5 fA
2 fA / 500 aA
b
1 nA 0 |I| 1.15 nA 50 fA 10 fA
10 nA 0 |I| 11.5 nA 500 fA 10 fA
100 nA 0 |I| 115 nA 5 pA 100 fA
1 μA 0 |I| 1.15 μA 50 pA 1 pA
10 μA 0 |I| 11.5 μA 500 pA 10 pA
100 μA 0 |I| 115 μA 5 nA 100 pA
1 mA
0 |I| 1.15 mA 50 nA 1 nA
10 mA 0 |I| 11.5 mA 500 nA 10 nA
100 mA 0 |I| 100 mA 5 μA 100 nA