Agilent B1500 Users Guide, Edition 7 10-75
Application Library and Utilities
Desktop EasyEXPERT
Calibration Differences on Calibration window
Calibration window provides Module Self Calibration screen only.
Module Self Calibration screen provides Enable Auto Calibration only.
Configuration,
Main Frame
Differences on Configuration window Main Frame screen
Line Frequency just displays the setting of 4155/4156.
Firmware Rev. displays as follows.
[HOSTC Rev.]:[SMUC Rev.]:[ADC Rev.]
Main Frame Diagnosis is not supplied.
Configuration,
Module
Differences on Configuration window Module screen
Self-test cannot be performed.
Status, SCUU, and Notes columns are not supplied.
Start Self Test and Recover Module buttons are not supplied.
Accessory Configuration is not supplied.
I/V Sweep Differences on I/V Sweep Classic Test
Time Stamp Name field is not supplied for 4155B/4156B.
SMU Pulse Period and Width value must satisfy Period Width + 4 ms.
I/V-t Sampling Differences on I/V-t Sampling Classic Test
Log sampling is not supported.
Sampling interval must be 60 μs to 65.535 s.
Number of samples must be 1 to 10001.
Base hold time must be 0 sec.
ADC and
Integration Time
Differences on A/D Converter & Integration Time Setup
ADC always shows HR ADC.
High Resolution ADC is set as shown in Table 10-17.
For the sampling measurement of Interval < 2 ms, the Integration Time setting is
ignored.