Agilent B1500 Users Guide, Edition 7 2- 35
Introduction
Specifications
Mainframe
The B1500A supports ten slots for plug-in modules.
Maximum module configuration
The B1500A can contain up to 4 dual-slot SMUs (HPSMUs) and 2 single-slot
SMUs (M
PSMUs and/or HRSMUs); it can contain up to 10 single-slot SMUs
(MPSMUs and/or HRSMUs); and it can contain any combination of dual-slot
and single-slot SMUs between these two extremes. Only one single-slot
MFCMU can be installed per B1500A mainframe. Up to five single-slot
HV-SPGUs can be installed per mainframe. Up to five single-slot WGFMUs can
be installed per mainframe.
Module
Slots
occupi
ed
Range of operatio
n
Measurement
re
solution
B1510A HPSMU 2 -200 V to 200 V,
-1A to 1 A
2 μV, 10 fA
B1511A MPSMU 1 -10
0 V to 100 V,
-100 mA to 100 mA
0.5 μV, 10 fA
B
1517A HR
SMU 1 -100 V to 100 V,
-100 mA to 100 mA
0.5 μV, 1 fA
E5
288A A
SU
a
a. The E5288A Atto Sense and Switch Unit (ASU) is connected with the
B1517A High Resolution SMU (HRSMU).
- -100 V to 100 V,
-100 mA to 100 mA
0.5 μV, 0.1 f
A
B1520A MFCMU 1 1 k
Hz to 5 MHz
0.035 fFrms
b
b. Dispersion of measurement values when connecting a DUT 10 pF to the
measurement terminals under the measurement condition of frequency
1 MHz, signal level 250 mVac, and measurement time 1 PLC. The dis-
play resolution is 0.000001 fF at 1 fF order by 6 digits display.
B1525A HV-SPGU 1 ± 40 V (80 Vp-p) 50 μV
B1530A WGFMU 1 PG Mode: -3 V to
3 V, -5 V
to 5 V
Fast IV M
ode: -3 V
to 3 V, -5 V
to 5 V,
0 V to 10 V, -10 V to
0 V
0.01
4 % of the
range
c
c. Display resolution