Agilent Technologies B1500A Guide de l'utilisateur Page 85

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 628
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 84
Agilent B1500 Users Guide, Edition 7 2- 37
Introduction
Specifications
MPSMU/HRSMU
MPSMU: Medium Power Source/Monitor Unit
HRSMU: High Resolution Source/Monitor Unit
Voltage range, resolution, and accuracy
See Table 2-15 for high resolution ADC and Table 2-17 for high speed ADC.
Current range, resolution, and accuracy
See Table 2-16 for high resolution ADC and Table 2-18 for high speed ADC.
Specified measurement resolution is limited by fundamental noise limits.
Mi
nimum dis
played resolution is 1 aA at 1 pA range by 6 digits. Measurements
made in the lower ranges can be greatly impacted by vibrations and shocks.
These specifications assume an environment free of these factors.
Power consumption
Voltage source mode:
Current source mode:
Voltage range (V)
Power consumption
a
(W)
a. Ic is the current compliance setting.
0.5 Ic × 20
2 Ic × 20
5 Ic × 20
20 Ic × 20
40 Ic × 40
100 Ic × 100
Voltage compliance
a
(V)
a. Vc is the voltage compliance setting.
Power consumption
b
(W)
b. Io is the output current.
Vc 20 Io × 20
20 < Vc 40 Io × 40
40 < Vc 100 Io × 100
Vue de la page 84
1 2 ... 80 81 82 83 84 85 86 87 88 89 90 ... 627 628

Commentaires sur ces manuels

Pas de commentaire