Agilent B1500 User’s Guide, Edition 7 9-11
If You Have a Problem
When You Perform Measurement
Unexpected Sampling Measurement Data is Returned
If sampling interval is set to a short time and if FILTER ON is set, you may get
unwanted data. FILTER ON causes a slower rise time, so short initial interval will
sample during this rise time.
To solve this problem:
• Set FILTER field to OFF if you set initial interval to a short time.
Some data may be skipped because measurement takes a long time. Meas
urement
takes a long time if measurement is performed in a low current range, if many
measurement channels are set up, or if analysis, such as moving a marker, is
performed during measurements.
To solve this problem:
• Measure current using a fixed range that is more than 10 μA. For measur
ement
ranges 10 μA or less, measurement takes longer than the specified integration
time.
• Do not perform analysis operation during measurement state.
MFCMU Causes Unbalance Condition
Extending measurement cables may cause the unbalance condition of the MFCMU.
To solve this problem:
• Perform the phase compensation. See “CMU Calibration” on page 4-37.
• The signal path from MFCMU to DUT must be as short as possible.
• Reduce stray capacitance and guard capacitance on the signal line.
Do not connect the Low (Lcur, Lpot) term
inal to the w
afer chuck. If the
MFCMU must be connected to the wafer chuck, use the High (Hcur, Hpot)
terminal to connect to the wafer chuck.