Agilent B1500 Users Guide, Edition 7 7- 31
Function Details
SPGU Module
Table 7-3 Output Timing for I/V-t Sampling, Simultaneous
Pulse output
conditions
So
urce Channel Operation
Starting During measurement After measurement Ending
Duration >
Meas
urement
time
SM
Us and
SPGUs start
the base value
output
simultaneously.
SMUs start
sampling
measurement
oper
ation.
After the base hold
tim
e, SMUs start th
e
source value output and
SPGUs start pulse
output simultaneously.
SMUs continue the source
v
alue output and SPGUs
continue pulse output until
the specified time elapses
or the repeat count reaches
the specified number.
After that, SMUs output
the b
ase or
source value set
by the Output Value after
Measurement, and SPGUs
output the base value.
If the bias
hold
functi
on is
ON, all
source
channels
continue
the output.
Else, all
s
ource
output
goes
to 0 V
in the
reverse
order of
the
Channel
Setup
definition.
Free run SMUs output the base or
s
ource value set b
y the
Output Value after
Measurement, and SPGUs
continue pulse output.
If the bias hold function is
OFF, SPGUs outp
ut the
base value.
Duration <
Measurement
time
A
fter the base hold
time, SMUs start the
source value output and
SPGUs start pulse
output simultaneously.
After the specified time
elapses or th
e repeat
count reaches the
specified number,
SPGUs output the base
value.
SMUs output the base or
s
ource value set b
y the
Output Value after
Measurement, and SPGUs
output the base value.