Agilent Technologies B1500A Guide de l'utilisateur Page 279

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 628
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 278
Agilent B1500 Users Guide, Edition 7 5- 11
Classic Test Definition
Multi Channel I/V Sweep
Measurement Setup
This setup screen is used to set the source channel outputs, the measurement ranging
mode, the sweep abort condition, the measurement channel A/D converter, the
series resistor, the filter, and the source output/measurement wait time.
VAR1 This area defines the primary sweep source output.
Direction Sweep direction, single (start to stop) or double (start to stop to
sta
r
t)
Li
ne
ar
/
Log Sweep output, LINEAR, LOG10, LOG25, or LOG50 where the
number foll
owing LOG indicates the number of sweep points in
one decade.
No of Step Number of sweep steps
Unit Source monitor unit (SMU) used for the VAR1 output
Name Variable name for the VAR1 output data
Start Sweep start value
Stop Sweep stop value
Compliance Compliance value. Enter the current limit value for the voltage
so
urce, or th
e voltage limit value for the current source.
Pwr Comp Power compliance value (in W), or OFF
Constant This area defines the constant output source.
Unit Source monitor unit (SMU) used for the constant source
V Name Variable name for the constant voltage output or measurement
data
I Name Va
riable name for the constant current output or measurement
data
Mode Ou
tput mode, V (voltage) or I (current)
Source Output value
Compliance Compliance value. Enter the current limit value for the voltage
so
urce, or th
e voltage limit value for the current source.
Vue de la page 278
1 2 ... 274 275 276 277 278 279 280 281 282 283 284 ... 627 628

Commentaires sur ces manuels

Pas de commentaire