Agilent B1500 Users Guide, Edition 7 2- 49
Introduction
Specifications
DC bias function
•DC bias
Range: 0 to ± 25 V
Resolution: 1 mV
Accuracy: ± (0.5 % + 5.0 mV) at the measurement port of the MFCMU or
t
he MFCMU cabl
e (1.5 m or 3.0 m)
Maximum DC bias current, supplemental information:
Output impedance: 50 Ω, typical
•DC bia
s monitor
Range: 0 to ± 25 V
Accuracy (open load): ± (0.2 % of reading + 10 mV) at the measurement
po
rt of th
e MFCMU or the MFCMU cable (1.5 m or 3.0 m)
Sweep characteristics
Available sweep parameters: DC bias voltage, oscillator level, frequency
Sweep type: linear, log
Sweep mode: single, double
Sweep direction: up, down
Number of measurement points: maximum 1001
Impedance range (Ω) Maximum DC bias current (A)
50
10
×10
-3
100
10
×10
-3
300
10
×10
-3
1×10
3
1×10
-3
3×10
3
1×10
-3
10×10
3
100×10
-6
30×10
3
100×10
-6
100×10
3
10×10
-6
300×10
3
10×10
-6