Agilent B1500 Users Guide, Edition 7 9-3
If You Have a Problem
When You Operate B1500A
When You Operate B1500A
This section covers the following basic problems that you may encounter when you
install or operate Agilent B1500A, and the solutions. See also “Error Codes” on
page 9-23.
“Power On Trouble”
“SCUU is not Detected”
“ASU is not Detected”
“External GPIB Devices are not Detected”
“To Simplify the Connections”
To set up the Windows, see “Inspection and Installation” on page 3-6
If you need to reinstall Agilent EasyEXPERT software, see “B1500A System
Recovery” on page 9-19
Power On Trouble
Connect the power cable to the B1500A firmly.
Connect the power cable to power outlet firmly.
Open the measurement terminals, and set the Standby swi
tch to the ON position.
NOTE Open the measurement terminals at the device side when turning the B1500A on.
Also disconnect the device from the measurement terminals and open the
measurement terminals after the measurement. If you leave the connection with the
device, the device may be damaged by unexpected operations or charge-up of
measurement cables.
SCUU is not Detected
If Agilent B1500A/EasyEXPERT cannot detect SCUU (SMU CMU unify unit), turn
B1500A off, check the cable connection, and reboot B1500A/EasyEXPERT.
For the installation and connection of SCUU
, see “To Connect SCUU/GSWU” on
page 3-21.