Agilent B1500 Users Guide, Edition 7 7- 27
Function Details
SPGU Module
PG Operation Mode
In the PG mode (pulse generator operation mode), the SPGU outputs 2-level or
3-level pulse voltage as shown in Figure 7-16.
The SPGU channels simultaneously start the pulse output and repeat the output
every pulse period
. The channels continue the pulse output until the specified time
elapses or the repeat count reaches the specified number of pulses.
For the SPGU pulse setup, see “SPGU Pulse Setup Parameters” on page 7-28.
For a summary of source output operation in sweep or sampling measurement mode,
see “Pulse O
utput Operation Summary” on page 7-30.
Figure 7-16 3-Level Pulse Output Examples
Pulse period
Width1
Width2
Delay2
Delay1
Output of
Additional Pulse
Output of
Primary Pulse
Pulse period
3-Level pulse output
Pulse period
Width1
Width2
Delay2
Delay1
Output of
Additional Pulse
Output of
Primary Pulse
Pulse period
3-Level pulse output
0 V
0 V
0 V
1 V
1 V
-1 V
-1 V
2 V
3 V
0 V
1 V
2 V
3 V
4 V