Agilent B1500 Users Guide, Edition 7 4- 67
Using EasyEXPERT
Data Display
Data Status
If measurement unit detects any status while measurement, the status code is
recorded with the measurement data. And the status code will be displayed in the
Data Display window as shown in the following example.
Example X-Y Graph Plot area
MARKER( 5.9000000 C 1.82520000 XV 3.33249E-3 )
This line will be displayed at the top of the plot area when marker is ON. The
marker is used t
o read data of the marker point.
List Display area
Index V1 V2 I2
123 5.9 V C 1.8252 V XV 3.33249 mA
Status Code SMU status code:
C: Compliance
T: Other channel compliance
X: Oscillation
•V: Overflow
MFCMU status code:
U: Unbalance
D: I/V amp saturation
•V: Overflow