Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1 9-3
If You Have a Problem
When You Operate B1500
When You Operate B1500
This section covers the following basic problems that you may encounter when you
install or operate the B1500, and the solutions.
• “Power On Trouble”
• “SCUU is not Detected”
• “ASU is not Detected”
• “N1265A/N1266A/N1268A is not Detected”
• “External GPIB Devices are not Detected”
• “To Simplify the Connections”
To set up the Windows, see Agilent B1500A or B1505A User’s Guide.
If you need to reinstall Agilent EasyEXPERT software, see “B1500 System
Recovery” on page 9-18
Power On Trouble
• Connect the power cable to the B1500 firmly.
• Connect the power cable to power outlet firmly.
• Open the measurement terminals, and set the Standby switch to the ON position.
NOTE Open the measurement terminals at the device side when turning the B1500 on. Also
disconnect the device from the measurement terminals and open the measurement
terminals after the measurement. If you leave the connection with the device, the
device may be damaged by unexpected operations or charge-up of measurement
cables.
SCUU is not Detected
If Agilent B1500A/EasyEXPERT cannot detect SCUU (SMU CMU unify unit), turn
the B1500A off, check the cable connection, and reboot the B1500A/EasyEXPERT.
For the installation and connection of SCUU, see Agilent B1500A User’s Guide.