Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1 6- 21
Remote Control Interface
RESult Subsystem
:RESult:FORMat
Sets or returns the output format of the test result data.
Syntax :RESult:FORMat mode
:RESult:FORMat?
Parameter mode Output format, TEXT|XTR. Data type is CPD.
TEXT: Text format, initial setting
XTR: EasyEXPERT test result data format
Query response mode<newline>
TEXT or XTR is returned. Data type is CRD.
Example :RES:FORM XTR
:RES:FORM?
:RESult:FORMat:ESCape
Sets or returns the newline character type used in the test result data.
Syntax :RESult:FORMat:ESCape mode
:RESult:FORMat:ESCape?
Parameter mode Type, 0|OFF|1|ON. Data type is Bool.
0 | OFF: ASCII code 0x0D (CR) and 0x0A (LF), initial setting
1 | ON: \r (CR) and \n (LF)
Query response mode<newline>
0 (OFF) or 1 (ON) is returned. Data type is Bool.
Example :RES:FORM:ESC ON
:RES:FORM:ESC?